Detection mode |
Fluorescence intensity – including FRET |
AlphaScreen ®/AlphaLISA® |
Luminescent (flashes and glows) – including FRET |
Time-resolved fluorescence |
TR-FRET |
Absorption spectrum UV/Vis |
Measurement mode |
Read above and below |
Endpoints and kinetics |
Sequential Multiple Stimulation |
Multi-sequential emission |
Scan spectrum (absorbance) |
Scale measurements |
Microplate form |
Plates 6 to 384, user definable |
LVis plate with 16 low-volume microspots (2 μL) |
Microplate holder |
Compatible with robots |
Light source |
High-energy xenon flash |
Probe |
Low noise photomultiplier tube |
Spectrophotometer CCD |
Wavelength selection |
Optical filters: excitation and emission rotary heads for 8 filters per side |
UV/vis absorption spectrometer: full spectrum or 8 separate wavelengths for <1 second/well |
Spectral band |
Filter |
240-740 nm or 240-900 nm for FI, FP, TRF |
240 – 740 nm for LUM |
Spectrometer |
220 – 1000 nm for ABS; wavelength accuracy: 0.5 nm |
Sensitivity |
FI |
<0.2 fmol/well FITC |
TRF |
<30 amol/well europium |
Premium TRF for Omega |
Premium TRF for Omega |
LUM |
<20 amol/well ATP |
AlphaScreen® |
5 pM (<100 amol/well P-Tyr-100, 384sv, 20 μL) |
ABS with spectrometer |
Selectable spectral resolutions: 1, 2, 5 and 10 nm |
OD range: 0 to 4 OD; photometric resolution: 0.001 OD |
Accuracy: <1% at 2 OD |
Accuracy: <0.5% at 1 OD and <0.8% at 2 OD |
Linearity: 0.8% at 2.0 OD |
Reading time |
Flight mode (1 flash) |
9 seconds (96), 16 seconds (384) |
Shake |
Linear, orbital and dual with user-defined time and speed |
Clean the vent hole |
System for pumping air or vacuum into the reader |
Annealing mode |
+4°C above ambient up to 45°C or 65°C |
The upper heating plate of the annealing chamber operates at a temperature 0.5°C higher than the lower plate. This prevents steam build-up on the lid or sealing gasket. |
Software |
Includes multi-user reader control software and MARS data analysis software |
Follow FDA 21 CFR Part 11 |
Size |
Width: 44cm, depth: 48cm, height: 30cm; Weight: 28 kg |
Optical filter |
Excitation and emission rotary head for 8 filters per side |
Optical path guide |
Above: liquid filled light guides |
Bottom: optical fiber |
Injection reagents |
Up to 2 reagent injectors built in |
Inject at measuring site (6 to 384 wells) |
Individual injection volumes per well (3 to 500L) |
Variable injection speed up to 420 L/s |
Up to four injections per well |
Reagents are flushed |