The ARYELLE 400 with a focal length of 400 mm and an aperture of f/10 features the highest spectral resolution of the wide-range spectrometer ARraY echELLE series. The patented optical setup with echelle grating, prism, imaging optics and entrance slit can individually be adapted to different customer applications. Depending on the required simultaneously detectable wavelength range, the ARYELLE 400 provides the measurement of two-dimensional echelle spectra. At standard spectrometer configuration with 50 µm entrance slit, a spectral resolving power of up to 30,000 for the VUV range (175-330 nm) or up to 15,000 for the UV-VIS-NIR range (330-850 nm) can be achieved. The optomechanical setup is designed to fit various CCD and ICCD cameras. Automatic recalibration of the wavelength scale is provided by a mercury lamp and an internal shutter.
The supplied operating software Sophi with optional LabVIEW library allows fully control of all spectrometer-detector functions via graphic user interface and provides automated measurement routines with an integrated scripting language. All lines of the gained spectrum are automatically analyzed with the integrated atomic spectra data base and qualitatively assigned to the corresponding elements. Quantitative analysis is based on calibration with reference materials.
Optical design |
echelle spectrograph |
Aperture |
f/10 |
Focal length |
400 mm |
Slit width |
50 µm |
Absolute accuracy |
spectral resolution / 4 |
Dynamic range |
15 bit, A/D conversion 16 bit |
Light coupling |
SMA fiber coupling or mirror optics |
Wavelength calibration |
with mercury lamp |
Computer |
PC incl. TFT or laptop with Microsoft Windows |
Software |
Sophi, LabVIEW library optional |
Dimensions without detector (L x W x H) |
(438 x 200 x 232) mm |
Weight without detector |
12 kg |
Setup |
1 |
2 |
3 |
4 |
5 |
6 |
Detector |
CCD |
CCD |
CCD |
CCD |
ICCD |
ICCD |
Wavelength range1 |
190-330 nm |
330-850 nm |
192-930 nm |
250-900 nm |
192-433 nm |
425-750 nm |
Spectral resolving power
λ/min. measurable FWHM1 |
30,000 |
15,000 |
15,000 |
45,000 |
14,000 |
20,000 |
Spectral resolution1,2 |
6-11 pm |
22-57 pm |
13-62 pm |
5.6-20 pm |
13-31 pm |
21-37 pm |
Detector resolution |
2,048 x 512 px |
2,048 x 512 px |
2,048 x 2,048 px |
2,048 x 2,048 px |
1,024 x 1,024 px |
1,024 x 1,024 px |
Image area |
27.6 x 6.9 mm² |
27.6 x 6.9 mm² |
27.6 x 27.6 mm² |
27.6 x 27.6 mm² |
13.3 x 13.3 mm² |
13.3 x 13.3 mm² |
Chopper |
yes |
yes |
yes |
yes |
no |
no |
Step width, min. |
0.1 µs |
0.1 µs |
0.1 µs |
0.1 µs |
1 ns |
1 ns |
Gate width |
– |
– |
– |
– |
5 ns |
5 ns |
1 Depending on the chosen grating and prism, other ranges possible (down to 175 nm)
2 Depending on the slit width, other widths possible
Benchtop sample chamber for spatially resolved LIBS analysis